Keeping connected medical devices safe

General information drawn from NHS England's guidance on connected medical devices, from GOV.UK and from legislation.gov.uk. It isn't medical, clinical or technical advice about any particular device. For a fault, start with the device's supplier and the Yellow Card scheme.

IoMT is short for the Internet of Medical Things. In a surgery, clinic, care home or hospital, it means medical equipment that connects to a network or the internet. Whoever looks after that network looks after them too.

The guidance quoted here is NHS England Digital's Guidance on protecting connected medical devices, last edited in November 2025. It was read on 26 November 2026.

Why they can't be patched like a laptop

A medical device is controlled more tightly than office equipment. The guidance says, "Modifications, including software and hardware, are not permitted, unless authorised by the medical device manufacturer, who will have conducted extensive validation and verification procedures in line with the appropriate regulations."

Updates wait for the maker's approval. In the guidance's words, "This can take 3 months (or longer) from the time that a security update is released."

Devices also stay in use for a long time. The guidance mentions "long device lifetimes (>10 years)", and devices still in use after their maker has stopped supporting them.

Knowing what is on the network

The first of the guidance's seven steps is an inventory. Its scope is "all medical devices that can receive or transfer patient data through a connection to a network or via the internet".

For each device, the guidance suggests recording the operating system, the software and its version, the ports and services it uses, and whether it holds or sends patient data. It says the maker may be able to supply a Manufacturer Disclosure Statement for Medical Device Security, known as an MDS2 form.

Keeping untrusted data away

The guidance lists the ways harmful data could reach a device:

It says, "We recommend that these routes be eliminated for all medical devices."

USB ports and disc drives need care. The guidance says that, because they may be part of normal or maintenance use, "they should never be altered without clear guidance and approval from the manufacturer of the specific medical device".

Limiting the damage if one is compromised

The guidance's answer is a separate part of the network for medical devices. It says, "Medical devices should be placed into network zones that minimise the traffic which can reach them."

Scanning tools need care. The guidance says, "Vulnerability management tools (such as port scanners) should be used with caution on areas of the network operating medical devices as they may disrupt the proper operation of the device and endanger patients."

It also suggests including a compromised medical device in regular incident response exercises.

Suppliers who connect in

The guidance notes that many manufacturers provide remote maintenance support. It says, "Any remote connections should only be initiated for specific tasks rather than allowing an 'always enabled' access policy."

It adds that any supplier connecting remotely should use multi-factor authentication (MFA).

Reviewing and retiring devices

The guidance treats review as a continual task.

A device's life expectancy should be part of the original purchase, it says. A device running software that is no longer supported should at least go on the organisation's risk register.

At the end, "All devices that store patient identifiable data must have all data forensically erased, so it is unrecoverable."

Two rules from outside the NHS guidance

The UK's security rules for consumer connected products carve out medical devices. Schedule 3 of the Product Security and Telecommunications Infrastructure Regulations 2023 excepts products that come under the Medical Devices Regulations 2002. A connectable product that runs medical-device software isn't excepted, though, so the security rules still apply to it.

Some alarms rely on an old mobile network. GOV.UK says, "UK mobile network operators will switch-off their 2G networks between 2029 and 2033." It lists medical pendants, or telecare alarms, among the equipment that can rely on 2G. GOV.UK suggests asking the service provider whether the equipment will work with 4G or 5G.

Reporting a problem

A device that injures someone, or almost does, can be reported through the Yellow Card scheme. GOV.UK says, "Anyone can report a problem." In Wales, in Scotland and in Northern Ireland, GOV.UK points to separate routes for device reports.

Device alerts and recalls come from the MHRA, the Medicines and Healthcare products Regulatory Agency, on GOV.UK.